![Rohde & Schwarz 1321.0615.02 User Manual Download Page 40](http://html1.mh-extra.com/html/rohde-and-schwarz/1321-0615-02/1321-0615-02_user-manual_1477771040.webp)
Getting Started
R&S
®
Spectrum Rider FPH
33
User Manual 1321.1011.02 ─ 11
For model 02, model 06 and model 13, N type connector is used. For model 26, RPC
3.5 mm connector is used, however with the installation of option R&S FPH-B100, N-
type connector is possible, see
Connect a cable or DUT to the RF input. Use a cable to connect the DUT to the R&S
Spectrum Rider, if necessary.
Make sure not to overload the R&S Spectrum Rider when a DUT is connected.
The maximum power that is permissible at the RF input is 20 dBm (or 100 mW).
The RF input is protected from static discharges and voltage pulses by a limiting cir-
cuit.
RF power overload
The R&S Spectrum Rider maybe loaded with up to 30 dBm (or 1 W ) for up to three
minutes. If you apply 1 W for a longer period, the R&S Spectrum Rider may be
destroyed.
Risk of electric shock
To avoid electrical shock the DC input voltage, you must never exceed the value speci-
fied on the housing.
Risk of damage of the R&S Spectrum Rider
To avoid damage to the coupling capacitor, input attenuator or the mixer, the DC input
voltage must never exceed the value specified in the data sheet.
3.2.2.2
BNC Connector
The BNC connector is located on the top of the R&S Spectrum Rider.
You can connect the BNC connector for various applications. It supports an external
trigger signal or an external reference signal.
Instrument Tour
www.allice.de
Allice Messtechnik GmbH