Remote Commands
R&S
®
Spectrum Rider FPH
335
User Manual 1321.1011.02 ─ 11
The numeric suffix at DISPlay is irrelevant for this command.
The numeric suffix at TRACe selects the number of the trace.
Suffix:
<1...2>
.
1...2
<1...2>
1...2
Parameters:
<DisplayMode>
WRITe | AVERage | MINHold | MAXHold | VIEW | FREeze |
INFinite
*RST:
WRITe
You can turn off the trace with
DISPlay<1...2>[:WINDow]:TRACe<1...2>[:STATe]
.
For more information, see chapter "Trace Mode".
Example:
SWE:CONT OFF
SWE:COUN 16
Turn on single sweep mode and sets the number of measure-
ments to 16.
DISP:TRAC:MODE MAXH
Activates MAXHold mode for the trace.
INIT;*WAI
Starts the measurement and waits for the end of the 16 sweeps.
FORMat:BORDer
<TransferOrder>
This command selects the format of binary data.
Parameters:
<TransferOrder>
NORMal | SWAPped
NORMal
The most significant byte is transferred first big endian).
SWAPped
The least significant byte is transferred first (little endian).
*RST:
SWAPped
Example:
FORM:BORD NORM
Changes the byte order to normal mode.
FORMat[:DATA]
<Format>[, <>]
This command selects the data format that is used for transmission of trace data from
the R&S Spectrum Rider to the controlling computer.
Note that the command has no effect for data that you send to the R&S Spectrum
Rider. The R&S Spectrum Rider automatically recognizes the data it receives, regard-
less of the format.
Parameters:
<Format>
ASCii | REAL