![Rohde & Schwarz 1321.0615.02 User Manual Download Page 328](http://html1.mh-extra.com/html/rohde-and-schwarz/1321-0615-02/1321-0615-02_user-manual_1477771328.webp)
Remote Commands
R&S
®
Spectrum Rider FPH
321
User Manual 1321.1011.02 ─ 11
The attenuation is coupled to the reference level. If you set the attenuation independ-
ently, the R&S Spectrum Rider turns off this coupling.
The R&S Spectrum Rider adjusts the reference level if it cannot be set for the current
RF attenuation.
Suffix:
<1...2>
.
1...2
RF port selection. Default selection at 1.
RF port 2 is only applicable for R&S ZNH.
Parameters:
<Attenuation>
Range:
0 dB to 40 dB
*RST:
0 dB (AUTO is ON)
Default unit: dB
Example:
INP:ATT1 30dB
Defines RF attenuation of 30 dB at RF port 1 and deactivates
coupling to the reference level.
INPut:ATTenuation:MODE
<AttenMode>
This command selects the attenuation mode.
Parameters:
<AttenMode>
LDIStortion | LNOise
LDIStortion
Selects "Auto Low Distortion" mode.
LNOise
Selects "Auto Low Noise" mode.
*RST:
LDIS
Example:
INP:ATT:MODE LNO
Sets the attenuation mode to Auto Low Noise.
INPut:ATTenuation:AUTO
<State>
This command couples and decouples input attenuation to the reference level.
Parameters:
<State>
ON | OFF
*RST:
ON
Example:
INP:ATT:AUTO ON
Couples the attenuation set on the attenuator to the reference
level.
INPut:GAIN:STATe
<State>
This command turns the preamplifier on and off.
Remote Control - Commands
www.allice.de
Allice Messtechnik GmbH