![Rohde & Schwarz 1321.0615.02 User Manual Download Page 455](http://html1.mh-extra.com/html/rohde-and-schwarz/1321-0615-02/1321-0615-02_user-manual_1477771455.webp)
Remote Commands
R&S
®
Spectrum Rider FPH
448
User Manual 1321.1011.02 ─ 11
Example:
CALC:DELT2:MAX:NEXT
Moves delta marker 2 to the next smaller maximum peak.
Usage:
Event
CALCulate<1...2>:DELTamarker<1...6>:MINimum[:PEAK]
This command positions a delta marker on the current trace minimum.
If necessary, the corresponding delta marker is activated first.
This command is an event and therefore has no *RST value and no query.
The numeric suffix at CALCulate is irrelevant for this command. The numeric suffix at
DELTamarker selects the deltamarker.
Example:
CALC:DELT3:MIN
Moves delta marker 3 to the trace minimum.
Usage:
Event
CALCulate<1...2>:DELTamarker<1...6>:X
<FrequencyOrTime>
This command positions a delta marker on a particular coordinate on the horizontal
axis.
Note that it is possible to place the marker outside the visible trace. In that case, this
value is invalid.
If necessary, the corresponding delta marker is activated first.
The numeric suffix at CALCulate is irrelevant for this command. The numeric suffix at
DELTamarker selects the deltamarker.
Parameters:
<FrequencyOrTime> Numeric value that indicates the coordinate on the horizontal
axis.
Range:
Maximum span.
Default unit: GHz
Example:
CALC:DELT:MOD REL
Delta marker positions are relative to marker 1.
CALC:DELT2:X 10.7MHz
Positions delta marker 2 10.7 MHz to the right of marker 1.
CALC:DELT2:X?
CALC:DELT2:X:REL?
Queries the absolute and relative position of delta marker 2.
CALCulate<1...2>:DELTamarker<1...6>:X:RELative
<FrequencyOrTime>
This command positions a delta marker on a position relative to the reference marker.
If necessary, the corresponding delta marker is activated first.
Remote Control - Commands
www.allice.de
Allice Messtechnik GmbH