![Rohde & Schwarz 1321.0615.02 User Manual Download Page 352](http://html1.mh-extra.com/html/rohde-and-schwarz/1321-0615-02/1321-0615-02_user-manual_1477771352.webp)
Remote Commands
R&S
®
Spectrum Rider FPH
345
User Manual 1321.1011.02 ─ 11
Parameters:
<State>
ON | OFF
*RST:
OFF
Example:
CALCulate<1...2>:MARKer<1...6>:X:
CALCulate<1...2>:MARKer<1...6>:X:SLIMits<1...2>:LEFT
<SearchLimit>
This command defines the left limit of the marker search range.
To use the command, you first have to turn on search limits with
CALCulate<1...2>:MARKer<1...6>:X:SLIMits<1...2>[:STATe]
The numeric suffix at CALCulate is irrelevant for this command.
The numeric suffix at MARKer selects the marker. The numeric suffix at SLIMits selects
the limit line.
Parameters:
<SearchLimit>
Sets the left marker search limit.
The unit in the spectrum analyzer mode depends on the mea-
surement, e.g. Hz for measurements in the frequency domain
and seconds for measurements in the time domain.
Range:
Maximum span.
*RST:
– (is set to the left diagram border when switching
on search limits)
Default unit: GHz
Example:
CALCulate<1...2>:MARKer<1...6>:X:
CALCulate<1...2>:MARKer<1...6>:X:SLIMits<1...2>:RIGHt
<SearchLimit>
This command defines the right limit of the marker search range.
To use the command, you first have to turn on search limits with
CALCulate<1...2>:MARKer<1...6>:X:SLIMits<1...2>[:STATe]
The numeric suffix at CALCulate is irrelevant for this command.
The numeric suffix at MARKer selects the marker. The numeric suffix at SLIMits selects
the limit line.
Remote Control - Commands
www.allice.de
Allice Messtechnik GmbH