![Rohde & Schwarz 1321.0615.02 User Manual Download Page 361](http://html1.mh-extra.com/html/rohde-and-schwarz/1321-0615-02/1321-0615-02_user-manual_1477771361.webp)
Remote Commands
R&S
®
Spectrum Rider FPH
354
User Manual 1321.1011.02 ─ 11
CALCulate<1...2>:LIMit<1...2>:DEFine
<Name>, <Description>, <X-unit>, <X-scale>,
<Y-unit>, <X0>, <Y0>, <X1>, <Y1>[, <X2>, <Y2>, <X3>, <Y3>, <X4>, <Y4>,
<X5>, <Y5>, <X6>, <Y6>, <X7>, <Y7>, <X8>, <Y8>, <X9>, <Y9>, <X10>,
<Y10>, <X11>, <Y11>, <X12>, <Y12>, <X13>, <Y13>, <X14>, <Y14>, <X15>,
<Y15>, <X16>, <Y16>, <X17>, <Y17>, <X18>, <Y18>, <X19>, <Y19>, <X20>,
<Y20>, <X21>, <Y21>, <X22>, <Y22>, <X23>, <Y23>, <X24>, <Y24>, <X25>,
<Y25>, <X26>, <Y26>, <X27>, <Y27>, <X28>, <Y28>, <X29>, <Y29>, <X30>,
<Y30>, <X31>, <Y31>, <X32>, <Y32>, <X33>, <Y33>, <X34>, <Y34>, <X35>,
<Y35>, <X36>, <Y36>, <X37>, <Y37>, <X38>, <Y38>, <X39>, <Y39>, <X40>,
<Y40>, <X41>, <Y41>, <X42>, <Y42>, <X43>, <Y43>, <X44>, <Y44>, <X45>,
<Y45>, <X46>, <Y46>, <X47>, <Y47>, <X48>, <Y48>, <X49>, <Y49>, <X50>,
<Y50>, <X51>, <Y51>, <X52>, <Y52>, <X53>, <Y53>, <X54>, <Y54>, <X55>,
<Y55>, <X56>, <Y56>, <X57>, <Y57>, <X58>, <Y58>, <X59>, <Y59>, <X60>,
<Y60>, <X61>, <Y61>, <X62>, <Y62>, <X63>, <Y63>, <X64>, <Y64>, <X65>,
<Y65>, <X66>, <Y66>, <X67>, <Y67>, <X68>, <Y68>, <X69>, <Y69>, <X70>,
<Y70>, <X71>, <Y71>, <X72>, <Y72>, <X73>, <Y73>, <X74>, <Y74>, <X75>,
<Y75>, <X76>, <Y76>, <X77>, <Y77>, <X78>, <Y78>, <X79>, <Y79>, <X80>,
<Y80>, <X81>, <Y81>, <X82>, <Y82>, <X83>, <Y83>, <X84>, <Y84>, <X85>,
<Y85>, <X86>, <Y86>, <X87>, <Y87>, <X88>, <Y88>, <X89>, <Y89>, <X90>,
<Y90>, <X91>, <Y91>, <X92>, <Y92>, <X93>, <Y93>, <X94>, <Y94>, <X95>,
<Y95>, <X96>, <Y96>, <X97>, <Y97>, <X98>, <Y98>, <X99>, <Y99>]
This command defines the shape of a limit line.
After you have defined the shape of the limit line, you still have to activate it with
CALCulate<1...2>:LIMit<1...2>:UPPer:SELect
effect.
The numeric suffix at CALCulate is irrelevant for this command. The numeric suffix at
LIMit selects the limit line.
Suffix:
<1...2>
.
1...2
<1...2>
1...2
Parameters:
<Name>
String containing the name of the limit line.
Note: if a limit line with the same name already exists, it is over-
written.
<Description>
String containing a comment for the limit line.
<X-unit>
HZ | S | M
Unit of the x-axis.
<X-scale>
ABSolute | RELative
Scale of the x-axis.
<Y-unit>
DB | DBM | DBMV | DBUV | DBUA | V | VOLT | W | WATT |
DBUVm | DBUam | V_M | W_M2 | VSWR
Unit of the y-axis.
Remote Control - Commands
www.allice.de
Allice Messtechnik GmbH