![Rohde & Schwarz 1321.0615.02 User Manual Download Page 95](http://html1.mh-extra.com/html/rohde-and-schwarz/1321-0615-02/1321-0615-02_user-manual_1477771095.webp)
Getting Started
R&S
®
Spectrum Rider FPH
88
User Manual 1321.1011.02 ─ 11
1 = Directional power sensor R&S FSH-Z14 or Z44
2 = Source
3 = Load
4 = USB binder adaptor (R&S FSH-Z144)
5 = USB port connector
Measuring the power
1. Press [MODE] key.
2. Select "Power Meter" softkey.
As soon as the R&S Spectrum Rider recognizes the power sensor, it displays the type
of the directional power sensor that is connected in the "Title bar" (see
).
After a few seconds, it also shows the forward power and return loss currently mea-
sured at the load.
Zeroing the power sensor
Before performing the power measurement, you should zeroing the power sensor.
After zeroing is done, the R&S Spectrum Rider displays the message "Power sensor
zero done" and again shows the power sensor softkey menu.
► Connect the R&S FSH-Z14 or R&S FSH-Z44 between the source and the load.
The R&S Spectrum Rider displays the measured forward power in dBm and the
SWR of the load. For more information on using the directional power sensor, see
Trying Out the Instrument
www.allice.de
Allice Messtechnik GmbH