![Rohde & Schwarz 1321.0615.02 User Manual Download Page 349](http://html1.mh-extra.com/html/rohde-and-schwarz/1321-0615-02/1321-0615-02_user-manual_1477771349.webp)
Remote Commands
R&S
®
Spectrum Rider FPH
342
User Manual 1321.1011.02 ─ 11
The numeric suffix at CALCulate is irrelevant for this command. The numeric suffix at
MARKer selects the marker.
Parameters:
<State>
ON | OFF
*RST:
OFF
Example:
CALC:MARK3 ON
Turns on marker 3.
CALCulate<1...2>:MARKer<1...6>:AOFF
This command turns off all active markers, delta markers and active marker measure-
ment functions.
This command is an event and therefore has no query and no *RST value.
The numeric suffix at CALCulate is irrelevant for this command. The numeric suffix at
MARKer selects the marker.
Example:
CALC:MARK:AOFF
Switches off all markers.
Usage:
Event
CALCulate<1...2>:MARKer<1...6>:FREQuency:MODE
<Mode>
This command selects the marker frequency display mode.
The numeric suffix at CALCulate is irrelevant for this command. The numeric suffix at
MARKer selects the marker.
Parameters:
<Mode>
FREQuency | CHANnel
FREQuency
Sets the marker frequency mode to frequency input (in Hz).
CHANnel
Sets the marker frequency mode to channel input (as a channel
number).
*RST:
FREQ
Example:
CALC:MARK:FREQ:MODE FREQ
Selects the frequency display mode.
CALCulate<1...2>:MARKer<1...6>:FUNCtion:CENTer
This command matches the center frequency to the frequency of a marker.
If you use a delta marker, the R&S Spectrum Rider turns it into a normal marker.
This command is an event and therefore has no *RST value and no query.
Remote Control - Commands
www.allice.de
Allice Messtechnik GmbH