Remote Commands
R&S
®
Spectrum Rider FPH
343
User Manual 1321.1011.02 ─ 11
The numeric suffix at CALCulate is irrelevant for this command. The numeric suffix at
MARKer selects the marker.
Example:
CALC:MARK2:FUNC:CENT
Matches the center frequency to the frequency of marker 2.
Usage:
Event
CALCulate<1...2>:MARKer<1...6>:FUNCtion:REFerence
This command matches the reference level to the power level of a marker.
If you use a delta marker, the R&S Spectrum Rider turns it into a normal marker.
This command is an event and therefore has no *RST value and no query.
The numeric suffix at CALCulate is irrelevant for this command. The numeric suffix at
MARKer selects the marker.
Example:
CALC:MARK1:FUNC:REF
Matches the reference level to the power level of marker 1.
Usage:
Event
CALCulate<1...2>:MARKer<1...6>:MAXimum[:PEAK]
This command positions a marker on the current trace maximum.
If necessary, the corresponding marker is activated first.
This command is an event and therefore has no *RST value and no query.
The numeric suffix at CALCulate is irrelevant for this command. The numeric suffix at
MARKer selects the marker.
Example:
CALC:MARK2:MAX
Moves marker 2 to the maximum peak.
Usage:
Event
CALCulate<1...2>:MARKer<1...6>:MAXimum:NEXT
This command positions a marker on the next smaller trace maximum.
If necessary, the corresponding marker is activated first.
This command is an event and therefore has no *RST value and no query.
The numeric suffix at CALCulate is irrelevant for this command. The numeric suffix at
MARKer selects the marker.
Example:
CALC:MARK2:MAX:NEXT
Moves marker 2 to the next smaller maximum peak.
Usage:
Event
Remote Control - Commands
www.allice.de
Allice Messtechnik GmbH