Remote Commands
R&S
®
Spectrum Rider FPH
325
User Manual 1321.1011.02 ─ 11
Example:
SOUR:TG:AUTO ON
SOUR:TG:FREQ:AUTO ON
Generates a CW signal with frequency coupled to the center fre-
quency of the current measurement.
SOURce:TG:FREQuency:OFFSet
<TgSourceFrequencyOffset>
This command defines a mathematical frequency offset for the tracking generator
Parameters:
TgSrcFreqOffset
Frequency offset of the TG source.
Range:
0 to 1.95
*RST:
0
Default unit: GHZ
Example:
SOUR:TG:FREQ:OFFS 1
Generates a tracking generator signal with an offset frequency
of 1 GHz.
SOURce:TG:POWer
<TgOutputLevel>
This command defines the absolute output level of the tracking generator.
Parameters:
<TgOutputLevel>
Numeric value in the range from 0 to -40 dBm.
*RST:
0
Default unit: dBm
Example:
SOUR:TG:POW -20
Defines an output level of -20 dBm.
SOURce:TG[:STATe]
<TgSourceState>
This command turns the tracking generator on and off.
Available when zero span has been turned on, for example with
.
Parameters:
<TgSourceState>
ON | OFF
*RST:
OFF
Example:
SOUR:TG:AUTO ON
SOUR:TG OFF
Stops the generation of a CW signal.
13.6.2.4
Setting the Bandwidth
The following commands configure the filter bandwidths of the R&S Spectrum Rider.
Note that both groups of commands (
BANDwidth
and
BWIDth
) are the same.
Remote Control - Commands
www.allice.de
Allice Messtechnik GmbH