![background image](http://html1.mh-extra.com/html/abbott/cell-dyn-3200/cell-dyn-3200_system-operators-manual_450985535.webp)
CELL-DYN
®
3200 System Operator’s Manual
11-5
9140181H—October 2001
Section 11
Quality Control
X-B File
Figure 11.3 X-B RBC Graphs Screen
The following soft key labels are displayed when the
[X-B FILE]
key is pressed:
X-B RBC DATA/GRAPHS
(Key label alternates between selections)
X-B WBC DATA/GRAPHS
(Key label alternates between selections)
PRINT
RETURN
For a description of how the X-B Program works, refer to
X-B Program Operation
later in this section.
Show Graphs
When the
[X-B RBC GRAPHS]
key or
[X-B WBC GRAPHS]
key is pressed, the
X-B
GRAPHS DISPLAY
screen for the appropriate parameter is displayed. (Refer to
Show Data
When the
[X-B RBC DATA]
key or
[X-B WBC DATA]
is pressed, the
X-B DATA DISPLAY
screen for the appropriate parameter is displayed. This screen is depicted in Figure
11.4. The screen displays the results for X-B batches 1–10 and the lower and upper
limits. The date and time that each batch was completed are also displayed. The
Page Down key on the keyboard is used to display batches 11–20. When batches
11–20 are displayed, the Page Up key on the keyboard is used to display batches
1–10.