CELL-DYN
®
3200 System Operator’s Manual
14-55
9140181H—October 2001
Section 14
Reticulocyte Package
4.
<Sex (M/F)>
This data entry field automatically displays the sex and
5.
<DOB>
birthdate
of the patient if it was found in the Standard
Hematology Data Log. If the information was not found
in the Standard Hematology Data Log, the sex and birth
date of the patient can be entered here.
6.
<Col>
This field is used to enter the collection date and time for
the specimen. (4 digits must be entered for the year)
7.
<Dr>
This data entry field automatically displays the name of
the patient’s doctor if it was found in the Standard
Hematology Data Log. If the information was not found
in the Standard Hematology Data Log, the name of the
patient’s doctor can be entered here (up to 22 characters).
8.
<Cmt>
This field is used to enter the comments. (Up to 16
characters may be entered.)
9.
<Limits>
This data entry field automatically displays the number of
the Limit Set that will be applied to the sample results.
NOTE:
The Limit Set applied to the reticulocyte sample may be changed
after the specimen has been run. Refer to the description for the
[EDIT SPECIMEN]
key given in
Reticulocyte Menu Options,
Reticulocyte Data Log Menu, Display Specimen Soft Key
within
this chapter.
These demographics can be entered or changed before the Reticulocyte specimen
is processed, or while the
EDIT SPECIMEN
screen in the Reticulocyte Data Log is
displayed.
Top Center
The Status Box is displayed in the top center of the
RETIC RUN RESULT
screen. This
box appears on every screen to show the following information:
• The menu in use (such as
RETIC RUN RESULT
).
• The status of the Analyzer (such as
Ready
,
Not ready
, and
FAULT
messages).
• Status and instructive messages. During the Reticulocyte Run cycle these are:
Aspirating
Remove Specimen
Dispensing
Rinsing
Ready