CELL-DYN
®
3200 System Operator’s Manual
9140181H—October 2001
Troubleshooting and Diagnostics
Tables for Instrument Conditions and Messages
Section 10
Loader Fault 149: mix zone rack position error
/BL
The tube position bar code number read was not the number expected.
– – – - Probable Cause – – –
– – – - Corrective Action –
1. The rack did not advance.
a. A mechanical problem is
preventing the rack from
advancing.
1. The rack did not advance.
a. Check for an obstruction that is
preventing the rack from
advancing and, if one is found,
remove it.
b. The rack and/or Sample Loader
Tray are dirty.
b. Remove the Loader Cover.
Using a DYN-A-WIPE pad or
lint-free wipe moistened with
deionized water, clean the rack
and Sample Loader Tray. Then,
reinstall the Loader Cover, reset
the racks, and reset the Sample
Loader by pressing the following
keys in order:
[CLEAR FAULT]
,
[START LOADER]
,
[RESET LOADER]
.
c. A failure of the Air Cylinder or
the Air Cylinder pressure system
has occurred.
c. If the message appears
repeatedly, contact your Abbott
Hematology Customer Support
Center.
Table 10.2 General Fault Conditions (Continued)