To Customize QC Log Display
1. Select a file from the
QC SET UP
menu
screen by moving the cursor to the
desired file.
2. Press
[CUSTOMIZE DISPLAY]
to display the
CUSTOMIZE QC DISPLAY
screen
for the selected file.
3. If necessary, press
[CUSTOM PLACEMENT]
to display the
CUSTOMIZE QC
DISPLAY
screen and the
[SELECT PARAMETER]
key.
4. Use the Arrow keys on the keyboard to move the cursor to the desired
parameter in the listing under the matrix.
5. Press
[SELECT PARAMETER]
. The selected parameter is highlighted and the
cursor moves to the first position in Group 1.
NOTE:
The key label changes to
[PLACE PARAMETER]
and a
[CANCEL
SELECTION]
key is displayed.
6. If necessary, use the Arrow keys on the keyboard to move the cursor to the
desired location and press
[PLACE PARAMETER]
.
NOTE:
When the
[PLACE PARAMETER]
key is pressed, the selected
parameter is displayed in the position indicated by the cursor and
the cursor is then advanced to the next parameter in the listing
under the matrix.
7. Repeat steps 4–6 until all selections have been made.
8. If desired, press the Print screen key on the keyboard to obtain a printout of
the selected groups.
9. Press
[RETURN]
to return to the
QC SET UP
menu
screen.
10. Repeat this procedure to customize the display for other QC logs.
Standard Groups
Predetermined groups of parameters, called Standard Groups, may be selected by
pressing the
[STANDARD GROUPS]
key (refer to Figure 5.14). When the
[STANDARD
GROUPS]
key is pressed, the following soft key labels are displayed:
WBC GROUP
RBC GROUP
PLT GROUP
DIFF GROUP
LATEX SET
CUSTOM PLACEMENT*
RETURN
* This key is used to return to the
CUSTOMIZE QC DISPLAY
screen for operator-
selected placement.