CELL-DYN
®
3200 System Operator’s Manual
14-41
9140181H—October 2001
Section 14
Reticulocyte Package
Fault Report
When the
[FAULTS REPORT]
key is pressed, information regarding any pending
fault is displayed on the screen. The screen displays the words
OPERATOR
CORRECTABLE FAULT REPORT:
or
FATAL FAULT REPORT
and any
additional information available. If there is no fault, the screen displays the words
NO FAULT PENDING
. Operator-correctable faults (for example, Waste Full,
Diluent Empty) can be cleared after the corrective action has been taken by
pressing the
[CLEAR FAULT]
key.
Execution Times
Display the time in tics used for various cycles during routine function of the
system.
NOTE:
A tic is an integer that equals 0.055 seconds.
Retic Count Rate
Figure 14.23 Reticulocyte Count Rate Screen (Tabular Format)
When the
[RETIC CNT RATE]
key is pressed followed by
[RETIC CNT DATA]
, the
following soft key labels are displayed:
RETIC CNT GRAPH
RETURN
RETIC
CNT GRAPH
DIAGNOSTICS MENU
Ready
Aug 02 2001
Operator ID
Retic Seq#
14:32
R0482
TS
RETURN
RETIC:
TOTAL COUNT: 31693
TIME:
0.50
1.00
1.50
2.00
2.50
3.00
3.50
4.01
COUNT:
2181
4204
6348
8458
10517
12705
14832
16936
RATE:
4362.00
4046.00
4288.00
4220.00
4118.00
4376.00
4254.00
4166.34
TIME:
4.51
5.01
5.51
6.02
6.52
7.02
7.50
COUNT:
19161
21247
23395
25507
27596
29692
31693
RATE:
4405.94
4172.00
4296.00
4182.18
4178.00
4192.00
4125.77