10-48
CELL-DYN
®
3200 System Operator’s Manual
9140181H—October 2001
Troubleshooting and Diagnostics
Tables for Instrument Conditions and Messages
Section 10
Runtime error
/BL
– – – - Probable Cause – – – –
– – – - Corrective Action – – –
1. An illegal software operation was
requested by the Analyzer.
1. Record what happened when the
message was displayed.
2. Initialize the Analyzer by pressing
the
[INITIALIZATION]
key on the
second
DIAGNOSTICS MENU
screen. Continue processing
samples. If the fault recurs, contact
your Abbott Hematology Customer
Support Center.
Bad checksum in non-volatile RAM
/BL
– – – - Probable Cause – – – –
– – – - Corrective Action – – –
1. When the system was powered ON,
the Analyzer did not transmit the
correct message to the Data Module.
1. Power OFF and ON again.
2. If the fault recurs, contact your
Abbott Hematology Customer
Support Center.
Bad monitor command
/BL
NOTE:
This message pertains to a software execution monitor on the Analyzer,
not the video display screen (CRT).
– – – - Probable Cause – – – –
– – – - Corrective Action – – –
1. During the initialization process, the
Data Module and Analyzer did not
communicate properly.
1. Initialize the Analyzer by pressing
the
[INITIALIZATION]
key on the
second
DIAGNOSTICS MENU
screen.
If the fault recurs, power OFF and
ON again. Contact your Abbott
Hematology Customer Support
Center if the message appears
repeatedly.
Table 10.2 General Fault Conditions (Continued)