CELL-DYN
®
3200 System Operator’s Manual
5-27
9140181H—October 2001
Section 5
Operating Instructions
Procedure: Load From Disk
1. Press
[QUALITY CONTROL]
to display a list of QC files.
2. Use the arrow keys on the keyboard to move the cursor to the desired file.
Type the file name (e.g., Low L0036) and press the Enter key on the keyboard
to save the name and advance the cursor to the next file.
NOTE:
The file must be empty in order to load the information from the
disk.
3. When the desired files have been named, use the Arrow keys on the keyboard
to move the cursor back to the first file desired for data entry.
4. Press
[QC LIMITS]
followed by
[MEANS/LIMITS]
or
[RANGE ENTRY]
to display the
QC MEANS/LIMITS ENTRY
or
RANGE ENTRY
screen for the selected file.
5. Press
[LOAD FROM DISK]
to display the
LOAD FROM DISK
screen.
6. Insert the disk containing the assay information for the relevant lot number
into the Data Station disk drive.
NOTE:
Be certain to carefully check lot numbers. Be sure the lot number
on the disk matches the lot number that is being put into use. If the
lot number is included in the filename, be sure the disk contains
assay information for that lot number.
7. Check the status box to be sure the correct file is selected and press the
appropriate soft key:
[LOAD LOW]
to load the low control assay data.
[LOAD NORMAL]
to load the normal control assay data.
[LOAD HIGH]
to load the high control assay data.
8. The limits are displayed for the selected file. If desired, the limits may be
edited.
9. Press
[RETURN]
to return to the
QC MENU
screen.
10. Select the next file and repeat steps 7 and 8 to load the assay data for the
appropriate level of control.
11. Repeat steps 8-11 until the assay data for each level has been loaded.
12. When all the assay data has been loaded, remove the disk from the disk drive.