9-46
CELL-DYN
®
3200 System Operator’s Manual
9140181K—July 2002
Service and Maintenance
Nonscheduled Maintenance Procedures
Section 9
Figure 9.18 HGB Flow Cell Access Tubing
Materials Required
1. Cleaning solution (1.25% sodium hypochlorite solution). See the formula for
mixing this solution under
Decontamination Procedures
earlier in this
section.
2. 10 mL syringe
3. Gloves, lab coat, and safety glasses
4. Deionized water
Procedure
1. In the
MAIN MENU
, press
[SPECIAL PROTOCOLS]
followed b
\>
DIS/ENABLE
ANALYZER]
and
[DISABLE ANALYZER]
to disable the Analyzer.
2. Gently pull on the hand hold of the right Front Cover to swing the cover open.
Pull the Tower Cover up and off the instrument (Tower Door on the CS model
must be open). Locate the HGB Flow Cell. (Refer to Figure 9.4.)
3. Remove the plastic drip pan under the Shear Valve by sliding it up and off the
support stand.
4. Fill the syringe with the bleach solution.
1 Solenoid 24
2 Tube Fitting
3 Shear Valve
1
2
3