Reticulocyte Package
Retic Menu Options
Section 14
Figure 14.18 Reticulocyte QC Log Screen
The
[RETIC QC]
key on the
RETIC MAIN
menu screen is used to display the
RETIC
QC LOG
screen. The Reticulocyte Package on the CELL-DYN 3200 System offers
six QC Logs with statistical and graphical analysis of the data. The statistical
analysis includes the mean, standard deviation, and coefficient of variation. The
results in each QC Log can be displayed as a Levey-Jennings chart. Westgard®
Rules can be applied to each QC Log. The rule options can be used independently
or in combination, at the operator’s discretion.
NOTE:
The
[RETIC QC LIMITS]
key and the
[QC FILE SET UP]
key are used to set
up the QC files.
The following soft keys are displayed on the
RETIC QC LOG
screen:
VIEW QC LOG
(Move the cursor with the arrow keys on the keyboard
to the QC file desired, then press this soft key.)
RETIC QC LIMITS
QC FILE SET UP
RETIC MAIN
VIEW
QC LOG
RETIC QC LOG
Ready
Feb 23 2001
Operator ID
Req Seq#
19:10
R0001
ACS
RETIC QC
QC FILE
SET UP
RETIC
LIMITS
1.
2.
3.
4.
5.
6.
File Name
Select a QC file with the arrow keys or enter a new file name.
FILE 1
FILE 2
FILE 3
FILE 4
FILE 5
FILE 6
0
0
0
0
0
0
# Specimens
MAIN