CELL-DYN
®
3200 System Operator’s Manual
6-31
9140181K—July 2002
Section 6
Calibration Procedures
b. If the value is highlighted in green, that parameter needs to be calibrated
and the new calibration factor displayed on the screen will replace the
existing calibration factor when the
[ACCEPT]
key is pressed. Green
indicates the Factor % Diff is within the range specified in column 3 of
Worksheet 6, Closed Mode Calibration Range Criteria, provided in
Manual Calibration Worksheets
at the end of this section.
c. If the value is highlighted in purple, it indicates the Factor % Diff is
greater than the value shown in column 4 of Worksheet 6, Closed Mode
Calibration Range Criteria, provided in
Manual Calibration Worksheets
at the end of this section.
d. If all Factor % Diff values are uncolored, go to
Calibration Not Required
below. If at least one Factor % Diff is green, go to
Some Parameters
Need Calibration
below. If at least one Factor % Diff is purple, go to
subsection
Over-Limit Parameters
below.
Calibration Not Required
1. If all the Factor % Diff values are uncolored, then calibration is not required
in Closed Mode. No update to the Calibration Log is required and no
calibration confirmation is required.
2. If Open Mode has not been calibrated, go to subsection
Auto-Cal — Open
Mode
, otherwise return to the
MAIN MENU
by pressing
[RETURN]
followed by
[MAIN]
.
Some Parameters Need Calibration
1. If at least one Factor % Diff value is highlighted in green and if no Factor %
Diff values are highlighted in purple, press
[ACCEPT]
followed by
[CONFIRM
ACCEPT]
to calibrate the Closed Mode. Go to subsection
Completing
Calibration Log
below.
Over-Limit Parameters
1. If the Factor % Diff for a parameter is highlighted in purple, it indicates there
may be an instrument problem. In this case, do the following:
a. Determine if any component on the instrument was changed. This could
affect calibration. Such components include the Shear Valve, Optical
Flow Cell, Hemoglobin Flow Cell, or one of the syringes.
b. If a component was changed, then treat the result as if it fell within the
“calibration range” (even though it is greater than the upper limit).
CALIBRATION IS REQUIRED for that parameter. Use the arrow keys
to move the cursor to the Factor % Diff value. Press the Enter key to
change the purple to green, indicating calibration is now required. Press
[ACCEPT]
followed by
[CONFIRM ACCEPT]
to calibrate this “over limit”
parameter along with any other “calibration required” parameters. Go to
subsection
Completing Calibration Log
below.