Section 10
Troubleshooting and Diagnostics
6. NOC:
• Follow the instructions given in step 5 for WBC, substituting the
[NOC CNT]
key for the
[WOC CNT]
key for displaying and printing rate and graph data.
Troubleshooting Imprecise or Inaccurate Data
– – – - Probable Cause – – – –
– – – - Corrective Action – – –
1. Improper Sample Mixing
1. Verify all samples are well-mixed prior to
analysis. (Refer to
.)
2. Dirty Syringe(s)
1. Inspect/Clean syringe as directed is
Nonscheduled Maintenance Procedures
.
3. Leaking Syringe(s)
1. Inspect syringe for evidence of leaks.
2. Verify syringe connectors are secure.
3. Remove/Clean /Replace syringe as directed in
.
4. Dirty Shear Valve
1. Clean shear valve as directed in
5. Dirty/Obstructed Y valve
1. Clean Y valve as directed in
6. Worn Sample Transfer Pump Tubing
1. Inspect/Replace sample transfer pump tubing as
.