CELL-DYN
®
3200 System Operator’s Manual
10-91
9140181H—October 2001
Section 10
Troubleshooting and Diagnostics
No screen display on the Display Monitor
– – – - Probable Cause – – – –
– – – - Corrective Action – – –
1. Display Monitor is not ON or is not
connected to power source.
1. Verify the Display Monitor LED is
ON. If not, check that the power
cord is securely connected to the
power outlet. Try a different power
source.
2. Display Monitor brightness control
is turned down.
2. Adjust the brightness control under
the front control panel of the Display
Monitor until the image is visible.
3. The Display Monitor interface
connector (VGA, 25 pin) is not
securely connected.
3. Locate the connector on the rear
panel of the Data Module, reseat the
cable and secure the connection
properly.
NOTE:
For the location of the
VGA connector, refer to
.
4. Defective Display Monitor or other
component.
4. Contact your Abbott Hematology
Customer Support Center.
Table 10.4 Non-Functional Fault Conditions (Continued)