CELL-DYN
®
3200 System Operator’s Manual
9-57
9140181H—October 2001
Section 9
Service and Maintenance
Salt deposits and reagent residue may clog the flow system if they are not removed
prior to a prolonged period of inactivity or shipment. In addition, the tubing should
be removed from behind the Peristaltic Pump and from all of the Normally Closed
Valves. Leaving this tubing in the pump and in the valves while the instrument is
inactive may cause it to crimp permanently. The Peristaltic Pump and Normally
Closed Valves on the Analyzer flow panel are shown in Figure 9.4.
NOTE:
When starting the system after 2 weeks or more of inactivity, it is
necessary to replace the Diluent/Sheath filter before priming the system.
Materials Required
1. Gloves, lab coat, and safety glasses
2. Phillips-head screwdriver (#2)
3. Disinfectant (cleaning solution); a solution containing 0.5% sodium
hypochlorite. See the formula for mixing this solution under
Decontamination Procedures
earlier in this section.
4. Container with approximately 500 mL of deionized water.
5. Four plastic bags
If the instrument will be shipped, the following are also needed:
6. Cardboard Disk Drive Protector
7. Shear valve dummy center section
Procedure
1. In the
MAIN MENU
, press
[SPECIAL PROTOCOLS]
followed by
[MORE]
and
[PREPARE SHIPPING]
2. Follow the instructions displayed on the screen.
3. Press the
[PREPARE SHIPPING]
key again to begin the rinse cycle.
NOTE:
The message
SPECIAL PROTOCOL CYCLE IN PROGRESS
is
displayed during the process.
4. When the process has finished, the message
SPECIAL PROTOCOL CYCLE
HAS COMPLETED
is displayed.
NOTE:
If the reagent lines were cleaned because contamination was
suspected and the operator intends to continue processing patient
samples, do not perform any of the remaining steps. Instead,
reconnect the reagent lines, go to the
RUN
screen, and run
background counts until the counts are acceptable.
5. Turn the main power switch OFF. Also turn OFF power to the Display
Monitor, printers, and any other devices attached to the instrument.
6. Open Left and Right Front Covers as follows. Gently pull on the cover using
hand hold and swing door open about 30-45 degrees. Lift door fully upward
from bottom area closest to door hinge bracket, then rotate door fully open so
lift and hold feature on upper door hinge mechanism can be engaged. Lower
door gently. Pull the Tower cover up and off the instrument (Tower Door on
the CS model must be open).
a. On the CS model, remove the four Phillips-head screws holding the Front
Skirt to the instrument and remove the skirt.