CELL-DYN
®
3200 System Operator’s Manual
6-37
9140181H—October 2001
Section 6
Calibration Procedures
Whole Blood Calibration Reference Values Worksheet
Date:
Technologist:
NOTE:
The WBC value obtained on the Reference Instrument should be used for
calibrating both the WOC and NOC parameters on the CELL-DYN 3200
Instrument.
Calibration Procedure — Open Mode
Follow the procedure below for calibrating the instrument in the Open Mode.
NOTE:
Use the same Calibrator or Fresh Whole Blood samples for calibrating
Open and Closed Modes.
Determining New Calibration Factors
1. If necessary, go to the
RUN
menu and press
[CHANGE SAMPLER]
to select the
Open Mode.
2. In the
MAIN MENU
, press
[CALIBRATION]
to display the main
CALIBRATION
menu.