10-56
CELL-DYN
®
3200 System Operator’s Manual
9140181H—October 2001
Troubleshooting and Diagnostics
Tables for Instrument Conditions and Messages
Section 10
Download flow sequence timeout
/BL
The allotted time for flow sequence downloading was exceeded.
– – – - Probable Cause – – –
– – – - Corrective Action –
1. The flow sequence did not compile,
possibly due to interruption by a
fatal fault, an improperly seated
Tower Cover or door, or a fluidics
problem.
1. Reboot the System.
NOTE:
If you wish to use the
Sample Loader, reset the
racks before rebooting the
System.
2. If the message appears repeatedly,
contact your Abbott Hematology
Customer Support Center.
Command transmission timeout
The Data Module did not receive acknowledgment of a command sent to the
Analyzer within the allotted time.
– – – - Probable Cause – – –
– – – - Corrective Action –
1. A failure in the Analyzer hardware,
possibly due to an unplugged or
damaged HSSL cable, has occurred.
1. Check the HSSL cable connections.
Then, initialize the Analyzer by
pressing the
[INITIALIZATION]
key
on the second
DIAGNOSTICS MENU
screen. If the message appears
again, reboot the System.
NOTE:
For the location of the
HSSL ports, refer to
NOTE:
If you wish to use the
Sample Loader, reset the
racks before initializing or
rebooting the System.
2. A software error has occurred.
2. If the message appears repeatedly,
contact your Abbott Hematology
Customer Support Center.
Table 10.2 General Fault Conditions (Continued)