CELL-DYN
®
3200 System Operator’s Manual
6-47
9140181H—October 2001
Section 6
Calibration Procedures
Entering New Calibration Factors — Closed Mode
1. In the
CALIBRATION
menu, press
[ENTER FACTOR]
.
2. Use the arrow keys to select the first factor in the Closed Sampler column to
be changed. Type in the New Calibration Factor calculated in
Calculating
New Closed Mode Calibration Factors
above. Press the Enter key to save the
factor and advance the cursor. If necessary, use the arrow keys to select the
next factor to be changed.
NOTE:
Prior to exiting the screen, the operator may press the
[RESTORE
FACTORS]
soft key to recall factors, stored on the Hard Disk,
corresponding to the current mode — Open or Closed.
[RESET ALL
TO 1.000]
is used to reset all factors displayed on the screen to
1.000.
3. When all new factors have been entered, press
[RETURN]
to display the
CALIBRATION LOG
screen with the new calibration factors.
NOTE:
If no changes are made in the
ENTER CALIBRATION FACTOR
screen,
pressing
[RETURN]
will bypass the
CALIBRATION LOG
screen and
display the main C
ALIBRATION
menu.
4. Complete the Calibration Log (adding comments) as required. If a printout of
the log is desired, press
[PRINT LOG]
.
5. Press
[RETURN]
twice to return to the main
CALIBRATION
menu.
6. Press
[PRINT]
to obtain a copy of the new Closed Sampler Calibration Factors.
7. Press
[MAIN]
to return to the
MAIN MENU
.