14-50
CELL-DYN
®
3200 System Operator’s Manual
9140181H—October 2001
Reticulocyte Package
Routine Operation
Section 14
NOTE:
A message describing the fault is displayed on the bulletin line. A list of
fault conditions and corrective action is given in
Troubleshooting and Diagnostics
.
Patient Specimen
The
[PATIENT SPECIMEN]
key on the
RETIC SPECIMEN TYPE
screen is used to
display the first
RETIC PATIENT SPECIMEN
screen. (See the following figure.) The
patient specimen ID (up to 12 alphanumeric characters) is entered here. Type the
information and press the Enter key on the keyboard to confirm the entry. The
Standard Hematology Data Log will then be searched for the entered specimen ID.
NOTE:
If the patient name is used for the specimen ID, the name must be typed
exactly as it was originally entered in the Standard Hematology Data Log.
NOTE:
Specimen IDs must match exactly and are case sensitive.
Figure 14.28 The First Reticulocyte Patient Specimen Screen
If the specimen ID is found, the second
RETIC PATIENT SPECIMEN
screen is
displayed. (See the following figure.) This screen displays the patient demographic
information and the most recent RBC value found in the Standard Hematology
Data Log.
NOTE:
To ensure optimal flagging, perform the CBC run within 8 hours prior to
the Retic run on the same analyzer using the same specimen.
This screen also displays the following question:
Use demographics and RBC value?-(Y/N).
RETIC SPECIMEN TYPE
Ready
Feb 23 2001
Operator ID
Retic Seq#
19:34
R0002
ACS
CANCEL
Press ENTER to continue if no Hematology Specimen result is available.
ENTER SPECIMEN ID: 123456 Press ENTER to confirm.