Loader Fault 48: tube stuck in position 4
/BL (Continued)
2. The tube position sensor is dirty.
2. Remove the Loader Cover. Using a
DYN-A-WIPE pad or lint-free
wipe moistened with lens cleaner or
deionized water, clean the tube
sensors (the tube position 4 sensor
is on the left). Dry the sensors
thoroughly. Then, reinstall the
Loader Cover, reset the racks, and
reset the Sample Loader.
NOTE:
For the location of the tube
sensors, refer to the figure
in
3. A failure of the sensor or related
electronics has occurred.
3. If the message appears repeatedly,
contact your Abbott Hematology
Customer Support Center.
Tower Fault 49: tower cover open
/BL
The circuit formed when the Tower Cover is in place has been broken.
– – – - Probable Cause – – –
– – – - Corrective Action –
1. The Tower Cover has been removed
or is not seated properly.
1. Reinstall or reset the Tower cover.
Make sure the cover is held in
position by the magnets on the
instrument frame. Then reset the
racks and reset the Sample Loader
by pressing the following keys in
order:
[CLEAR FAULT]
,
[START LOADER]
,
[RESET LOADER]
.
2. A failure of the sensor or related
electronics has occurred.
2. Reset the racks and reset the
Sample Loader. If the message
appears repeatedly, contact your
Abbott Hematology Customer
Support Center.
Table 10.2 General Fault Conditions (Continued)