CELL-DYN
®
3200 System Operator’s Manual
9-39
9140181H—October 2001
Section 9
Service and Maintenance
Unclogging the Closed Sample Aspiration Needle
If a blockage is suspected, it may be cleared using the procedure described above
in
Closed Sample Needle Interior Cleaning
.
Syringe Cleaning
There are two syringe assemblies, each containing two syringes, on the Flow Panel
of the CELL-DYN 3200 system. The syringe assemblies are depicted in Figure
9.13.
Syringes on the CELL-DYN 3200 System should be cleaned only as necessary one
at a time to ensure that each syringe is replaced in the correct position. Replace
each syringe after it is cleaned and then remove the next one to be cleaned.
Figure 9.13 Syringe Assemblies
1 Sample Injection
Syringe
2 HGB Lyse Syringe
3 WBC Lyse Syringe
4 Diluent/Sheath
Syringe
5 Holding Blocks
1
2
3
4
5