![Abbott CELL-DYN 3200 System Operator'S Manual Download Page 169](http://html1.mh-extra.com/html/abbott/cell-dyn-3200/cell-dyn-3200_system-operators-manual_450985169.webp)
CELL-DYN
®
3200 System Operator’s Manual
5-17
9140181H—October 2001
Section 5
Operating Instructions
Patient age is entered in Weeks and/or Years. The range for Weeks is zero to 51.
The range for Years is 1 to 199. Any number input shall be interpreted by the
system as “equal to or less than.” Age ranges are automatically established by using
successively higher numbers. For example, if the number 1 is entered in the Years
field for Limit Set 2 and the number 5 is entered in the Years field for Limit Set 3,
then any patient whose age is 1 year or less will automatically be assigned Limit
Set 2 and any patient whose age is greater than 1 year but equal to or less than 5
years will automatically be assigned Limit Set 3.
The appropriate Limit Set is automatically assigned when a patient’s sex and date
of birth are entered on the
RUN
screen or Work List.
To Set Patient Limits
1. Press
[PATIENT LIMITS]
to display the
PATIENT LIMITS
screen.
A Patient Limit Set is displayed on the screen. When a Limit Set is displayed
on the screen, the set number (Limit Set 1, Limit Set 2, etc.) is displayed in the
status box and the soft key corresponding to that number is blanked out. The
other limit sets may be selected by pressing the appropriate soft key.
2. Select a Limit Set by pressing the appropriate
[LIMIT SET]
soft key.
3. To pre-assign a Limit Set based on a patient’s sex and age, enter one of the
three Sex codes and press the Enter key. The cursor moves to the Years field.
Enter a number for years and press Enter or use the arrow keys to move the
cursor to the Weeks field. Enter a number for Weeks and press Enter or use
the arrow keys to move the cursor to the desired limit entry field.
NOTE:
Limit Set 1 is the default Limit Set and cannot be altered.
4. With the cursor in the first entry field to be changed, type the desired number.
5. Press the Enter key on the keyboard to save the entry and automatically
advance the cursor to the next entry position.
6. Repeat steps 4 and 5 until all desired entries have been made.
7. If desired, press
[PRINT]
to obtain a printout of the Limit Set.
NOTE:
Retaining a hard copy of each Limit Set is recommended, as the
screens do not display names or categories for the limit sets.
8. Press the appropriate soft key to select another Limit Set and repeat steps
3–7 to pre-assign a Limit Set and to enter the desired limits.
9. Press
[RETURN]
to return to the
SET UP MENU
screen.