![background image](http://html1.mh-extra.com/html/abbott/cell-dyn-3200/cell-dyn-3200_system-operators-manual_450985186.webp)
5-34
CELL-DYN
®
3200 System Operator’s Manual
9140181H—October 2001
Operating Instructions
Set Up Instructions
Section 5
Customize QC Log Display (Standard Groups)
1. Select a file from the
QC SET UP
menu
screen by moving the cursor to the
desired file.
2. Press
[CUSTOMIZE DISPLAY]
to display the
CUSTOMIZE QC DISPLAY
screen
for the selected file.
3. Press
[STANDARD GROUPS]
to display the
CUSTOMIZE QC DISPLAY
screen
and key labels for Standard Groups.
4. Use the Arrow keys on the keyboard to move the cursor to the desired group
(1–4) location.
NOTE:
This number indicates the order in which the group of parameters
will be displayed (Group 1 on the first screen, Group 2 on the
second, etc.).
5. Press the soft key corresponding to the desired parameter group. This group
is displayed in the position indicated by the cursor.
6. Repeat steps 4 and 5 until all desired groups have been selected.
7. If desired, press the Print screen key on the keyboard to obtain a printout of
the configuration.
8. Press
[RETURN]
to return to the
QC SET UP
menu
screen.
9. Repeat this procedure to select standard groups for other QC logs.
Latex Set
This key is intended to be used by Abbott Service personnel only. If the
[LATEX
SET]
key is pressed, parameters in all groups will be affected. The
[LATEX SET]
key
is available from the
CUSTOMIZE QC DISPLAY
screen for Standard Groups. This
key is used to customize the QC file to store information generated by latex
particles. Information is stored for each of the four angles of scatter which are used
to determine the differential.