CELL-DYN
®
3200 System Operator’s Manual
10-35
9140181H—October 2001
Section 10
Troubleshooting and Diagnostics
Clearing fault
/SB
– – – – – – – – – – – – – – – – – Explanation/Action– – – – – – – – – – – – – – – –
The
[CLEAR FAULT]
key was pressed after an operator-correctable fault was
detected.
Resume operation when READY is displayed in the Status Box and the READY
indicator light on the status indicator panel is illuminated in green.
Limits were changed to correct out-of-range values
/BL
– – – – – – – – – – – – – – – – – Explanation/Action– – – – – – – – – – – – – – – –
A mathematically incorrect limit was manually entered (using
[MEANS/LIMITS]
)
during setup of a QC file. The numbers entered generated a range containing a
number greater than the largest number allowed or less than zero. Therefore, the
limits were automatically changed.
Check to be sure the entered values are correct. If appropriate, recalculate the
mean and limits and enter correct values.
Entries making upper limit = lower limit were rejected
/BL
– – – – – – – – – – – – – – – – – Explanation/Action– – – – – – – – – – – – – – – –
A mathematically incorrect limit was entered (using
[RANGE ENTRY]
) during
setup of a QC file. The currently entered numbers are not accepted and the
previously entered numbers for the parameter(s) remain.
Check to be sure the entered values are correct.
Limits were exchanged to make upper > lower
/BL
– – – – – – – – – – – – – – – – – Explanation/Action– – – – – – – – – – – – – – – –
A mathematically incorrect limit was entered (using
[RANGE ENTRY]
) during
setup of a QC file. The numbers entered caused the upper limit to be less than the
lower limit. Therefore, the numbers were automatically exchanged.
Check to be sure the entered values are correct. If appropriate, enter correct
values.
Westgard Warning — See Levey Jennings
/BL
The message is displayed on the
VIEW QC LOG
screen only.
– – – – – – – – – – – – – – – – – Explanation/Action– – – – – – – – – – – – – – – –
The Westgard
®
Rules were selected during set up of the QC file and the data in
the file has violated one or more of the selected rules.
Review the data in the QC file and take appropriate action.
Table 10.1 Status
Conditions (Continued)