10-26
CELL-DYN
®
3200 System Operator’s Manual
9140181H—October 2001
Troubleshooting and Diagnostics
Troubleshooting Procedures
Section 10
6. Configure the
RUN
screen to display the appropriate graphic for the
parameter(s) for which the background exceeds the specifications:
Parameter
Appropriate Graphic
WBC
WBC histogram/scatterplots
RBC
RBC and PLT histograms/scatterplots
PLT
RBC and PLT histograms/scatterplots
HGB
WBC, RBC and PLT histograms
NOC
NOC histogram
Obtain several printouts of this information by running multiple background
cycles.
NOTE:
Instructions for customizing the RUN
screen display are given in
Set Up Instructions,
Customize Report, Customize Displayed Report
.
7. Configure the Data Log to display values for WBC, RBC, PLT, HGB, and
NOC. Obtain printouts of the Data Log, including the sequence numbers of
the background cycles.
NOTE:
Instructions for customizing the display and printout of the Data
Log are given in
Subsection:
Using the Data Log, Data Log Set Up Procedures
.
Troubleshooting Reagent Problems
If a reagent (or reagents) is suspected as the cause of a particular problem, replace
the container. However, the Analyzer has reservoirs that contain a small amount of
reagent to maintain the supply within the system. This supply must be depleted
before installing the new reagent.
NOTE:
There is no reservoir for either the HGB Lyse reagent or WBC lyse
reagent. The amount of lyse contained in the lyse supply tubing is
sufficient to maintain the system’s supply. The lyse tubing is drained and
filled by pressing the appropriate reagent key displayed on the
REAGENT
RESERVOIR
screen described below.
To ensure that the new reagent is actually in the system, proceed as follows:
1. From the first
SPECIAL PROTOCOLS
screen, press
[REAGENT RESERVOIR]
.
2. From the
REAGENT RESERVOIR
screen, press the soft key for the desired
reagent and follow the instructions given on the screen.
3. Wipe the reagent line with a lint-free wipe before placing it in the new
container. Place the line in the container and secure the cap.
4. Press the appropriate soft key to refill the reservoir.
5. Run five Background counts before assessing the results.