![Abbott CELL-DYN 3200 System Operator'S Manual Download Page 462](http://html1.mh-extra.com/html/abbott/cell-dyn-3200/cell-dyn-3200_system-operators-manual_450985462.webp)
10-28
CELL-DYN
®
3200 System Operator’s Manual
9140181K—July 2002
Troubleshooting and Diagnostics
Troubleshooting Procedures
Section 10
4. WBC:
• Configure the
RUN
screen to display the following and obtain printouts by
pressing the Print screen key on the keyboard. (This requires two printouts
per sample.)
Printout 1:
Size-Cmp (0—10)
Grn-Lob (90D—90)
WBC 0—90 deg
RBC 90—10 deg
• Printout 2:
N-L-M histogram
M-P histogram
NOC histogram
If necessary, refer to the directions given in
for customizing the display.
• Obtain printouts of the kinetic data for WBC for several samples. From
the first
DIAGNOSTICS MENU
screen, press
[CNT RATE SUMMARY]
. Press
[WOC CNT RATE]
to display the data for the kinetic rate followed by
[PRINT]
to obtain a printout. Press
[WOC CNT GRAPH]
to display the graph
of the kinetic data followed by
[PRINT]
to obtain a printout.
• If there is a flagging problem or a problem with an abnormal sample,
obtain a printout of a normal sample for comparison.
• Obtain a printout of the
RAW DATA SUMMARY
screen immediately after
the problem sample is run. From the first
DIAGNOSTICS MENU
screen,
press
[RAW DATA SUMMARY]
followed by
[PRINT]
to obtain a printout.
• Configure the Data Log to display and print WBC values. (If necessary,
refer to the instructions given in
the results for the last 100 cycles.
5. RBC, HGB, MCV and PLT:
• Configure the
RUN
screen to display the RBC and PLT
histograms/scatterplots. Obtain printouts of problem samples.
• Obtain a printout of the
RAW DATA SUMMARY
screen immediately after
the problem sample is run. From the first
DIAGNOSTICS MENU
screen,
press
[RAW DATA SUMMARY]
followed by
[PRINT]
to obtain a printout.
• Obtain a printout of the X-B data. In the
MAIN MENU
, press
[QUALITY
CONTROL]
followed by
[X-B FILE]
. If necessary, press
[X-B RBC DATA]
to
display the X-B RBC data and press
[PRINT]
to obtain a printout.
• Check the Hemoglobin Reference and verify that its values are within
2050 ±
250.
• Check the hemoglobin Flow Cell if the reference value is <1800.
• If the Hemoglobin Reference Value is >2300, call your Hemotology
Customer Support Center.