9-6
CELL-DYN
®
3200 System Operator’s Manual
9140181H—October 2001
Service and Maintenance
Special Protocols
Section 9
Pressing the
[DISABLE ANALYZER]
key prevents the Analyzer from cycling while
certain maintenance procedures are performed. After the Analyzer has been
disabled, the key toggles to
[ENABLE ANALYZER]
.
When the
[ENABLE ANALYZER]
key is pressed, the Analyzer is returned to the state
prior to disabling.
Figure 9.3 Second Special Protocols Screen
More
When the
[MORE]
key is pressed, the second
SPECIAL PROTOCOLS
screen and the
following soft key labels are displayed, as shown in Figure 9.3:
AUTO CLEAN
DAILY SHUTDOWN
PREPARE SHIPPING
CLEAN NEEDLE
EXTENDED AUTOCLEAN
MORE
MAIN
Auto-Clean
NOTE:
This procedure is performed only in the Open Mode. If necessary, press
[CHANGE SAMPLER]
in the
RUN
screen to select Open Mode.