9-62
CELL-DYN
®
3200 System Operator’s Manual
9140181H—October 2001
Service and Maintenance
Nonscheduled Maintenance Procedures
Section 9
19. Run at least three background counts. Check that the background counts are
acceptable before running controls or patient samples.
20. If the counts are unacceptable, troubleshoot accordingly (refer to
Troubleshooting and Diagnostics
Rerun samples and confirm that problem is resolved.
21. Record the completion of this procedure in the maintenance log.