![Abbott CELL-DYN 3200 System Operator'S Manual Download Page 183](http://html1.mh-extra.com/html/abbott/cell-dyn-3200/cell-dyn-3200_system-operators-manual_450985183.webp)
CELL-DYN
®
3200 System Operator’s Manual
5-31
9140181H—October 2001
Section 5
Operating Instructions
To Enter Replicate ID Data
1. If necessary, from the
QC FILE SET UP
screen, press
[REPLICATE ID]
to display
the <
Replicate Id
!
entry field.
2. The cursor is in the <Replicate ID> entry field. Type the sample ID number
(up to 12 alphanumeric characters may be entered) and press the Enter key
on the keyboard to save the entry and advance the cursor to the
<Westgard Rule Selection> entry fields.
3. Select the Westgard
®
Rules as directed in the To Enter Lot Number Data
procedure.
Customize Display
The
[CUSTOMIZE DISPLAY]
key in the
QC SET UP
menu is used to customize the
display of information in the QC logs. Figure 5.13 shows the
CUSTOMIZE QC
DISPLAY
screen with the standard groups displayed. Figure 5.14 shows the
CUSTOMIZE QC DISPLAY
screen where the WBC and RBC groups have been
customized.
The
CUSTOMIZE QC DISPLAY
screen and the following key labels are displayed
when the
[CUSTOMIZE DISPLAY]
key is pressed:
SELECT PARAMETER
STANDARD GROUPS
RETURN
The screen displays a matrix of 4 rows of parameters that are currently selected in
Groups 1 through 4. A list of all available parameters is displayed under the matrix.
The standard groups, which include predetermined parameter sets, are identified as
follows:
Group 1
WBC parameters
Group 2
RBC parameters
Group 3
PLT parameters
Group 4
WBC Diff parameters
On the
CUSTOMIZE QC DISPLAY
screen, Parameter Group 1 is displayed (in the
order indicated from left to right) on the first
VIEW QC LOG
screen. The remaining
groups are displayed on subsequent screens that are accessed by pressing the Right
Arrow key on the keyboard. The Left Arrow key is used to page back through the
screens to the first screen.
The
[STANDARD GROUPS]
key is used to select a predetermined group of
parameters that will be placed on a designated page. The display may be
customized by selecting the individual parameters, standard groups of parameters,
or a combination of the two.