10-14
CELL-DYN
®
3200 System Operator’s Manual
9140181H—October 2001
Troubleshooting and Diagnostics
Diagnostics Menu
Section 10
Figure 10.13 Vacuum Test Screen
Vacuum Test
When the
[VACUUM TEST]
key is pressed, the system releases the vacuum to
atmosphere and then determines the amount of time required for it to return to the
correct level. The key labels disappear and the incrementing time is displayed on
the screen. (See Figure 10.13.) When the test is finished, the time stops
incrementing and the key labels are displayed.
Pressure Test
When the
[PRESSURE TEST]
key is pressed, the system releases the pressure to
atmosphere and then monitors the amount of time required for it to return to the
correct level. The key labels disappear and the incrementing time is displayed on
the screen. When the test is finished, the time stops incrementing and the key labels
are displayed.