14-16
CELL-DYN
®
3200 System Operator’s Manual
9140181H—October 2001
Reticulocyte Package
Retic Menu Options
Section 14
Figure 14.5 Reticulocyte Patient Limits Screen
The
[RETIC PT LIMITS]
key is used to enter upper and lower flagging limits for
groups of patient samples. (For example, limits may be entered for adult males,
adult females, neonates, etc.)
Six sets of limits may be entered. Whenever a result falls outside an entered limit,
the result is displayed in color on the screen to alert the operator. Results displayed
in yellow are below the limit and results displayed in purple are above the limit.
The flagged result is underlined on the graphics report and in the Reticulocyte Data
Log when printed. Patient results that exceed the linearity specifications will be
suppressed and chevrons (
>>>>
) will be displayed and printed.
Procedure: Enter Retic Patient Limits
1. From the
RETIC SET UP
screen, press the
[RETIC PT LIMITS]
key to display the
RETIC PATIENT LIMITS
screen. A patient limit set is displayed on the screen.
The other five limit sets (Limit Set 2, Limit Set 3, etc.) may be displayed by
pressing the appropriate soft keys.
2. Use the arrow keys on the keyboard to move the cursor to the desired limit
entry field and type the desired number. Press the Enter key on the keyboard
to save the entry.
3. Repeat step 2 until all desired entries have been made.
4. If desired, press the
[PRINT]
key to obtain a printout of the Limit Set being
displayed.
NOTE:
Retaining a hard copy of each Limit Set is recommended, as the
screens do not display names or categories for the Limit Sets.
LIMIT SET 1
Ready
Feb 23 2001
Operator ID
Req Seq#
18:36
R-1
ACS
LIMIT
LIMIT
SET 3
LIMIT
SET 4
LIMIT
SET 5
RETURN
RETIC%
0.00
%
35.0
%
RETIC ABS
0.
K/uL
1500
K/uL
RBC
0.00
M/uL
9.99
M/uL
Lower Limits
Upper Limits
SET 2
LIMIT
SET 6