9-38
CELL-DYN
®
3200 System Operator’s Manual
9140181H—October 2001
Service and Maintenance
Nonscheduled Maintenance Procedures
Section 9
10. Place the needle back into the Wash Block, making sure the shorter part (vent
needle) faces the instrument, and place the flange into its slot in the top
bracket.
11. Reinstall the clip over the top of the needle and tighten the thumb screw.
12. Attach the vent tubing to the vent section of the needle and the aspiration
tubing to the aspiration section.
NOTE:
It is important that the tubing be attached correctly. (Refer to
Figure 9.12.)
NOTE:
Wetting the top portion of the needle will allow the tubing to slide
more easily.
13. Reattach the Tower Cover (Tower Door on the CS model must be open) and
press into place.
14. Press
[ENABLE ANALYZER]
followed by
[RETURN]
and
[MAIN]
to return to the
MAIN MENU
.
15. Run at least three background counts to rinse the system. Check that the
background counts are acceptable before running controls or patient samples.
If the counts are unacceptable, troubleshoot accordingly (refer to
Troubleshooting and Diagnostics
16. Record this maintenance in your maintenance log.