CELL-DYN
®
3200 System Operator’s Manual
10-49
9140181H—October 2001
Section 10
Troubleshooting and Diagnostics
Message retransmission failure
/BL
Communication from the Data Module to the Analyzer failed.
– – – - Probable Cause – – – –
– – – - Corrective Action – – –
1. A hardware failure or system error
has occurred.
1. Initialize the Analyzer by pressing
the
[INITIALIZATION]
key on the
second
DIAGNOSTICS MENU
screen. If the message appears
again, reboot the System.
NOTE:
If you wish to use the
Sample Loader, reset the
racks before initializing or
rebooting the System.
2. If the message appears repeatedly,
contact your Abbott Hematology
Customer Support Center.