5-28
CELL-DYN
®
3200 System Operator’s Manual
9140181H—October 2001
Operating Instructions
Set Up Instructions
Section 5
Set Up QC File
The
[SET UP QC FILE]
key is used to configure the QC file. If the file is used for a
commercial control, the lot number and expiration date may be entered by pressing
the
[LOT NUMBER]
key. If the file is used for a patient control, the ID number of the
control may be entered by pressing the
[REPLICATE ID]
key.
NOTE:
The system defaults to the
LOT NUMBER
screen upon initial set up.
Otherwise, the screen last selected will be displayed.
The
QC SET UP
screen is also used to select the Westgard
®
Rules that will be
applied to the QC data stored in the file. The following soft key labels are displayed
when the
[SET UP QC FILE]
key is pressed:
REPLICATE ID/LOT NUMBER
(The key alternates between the selections.)
TOGGLE ON/OFF
(This key is only present when the cursor is in one
of the Westgard
®
Rule Selection fields.)
PRINT
RETURN
The
LOT NUMBER ENTRY
and
REPLICATE ID ENTRY
screens are shown in Figures
5.11 and 5.12. The options available are as follows:
1. <Lot Number> and <Expiration Date> entry fields
These entry fields are displayed when the
[LOT NUMBER]
key is pressed. This
designation is intended for QC Files that are used for commercial controls.
2. <Replicate ID> entry field
This entry field is displayed when the
[REPLICATE ID]
key is pressed. This
designation is intended for QC Files that are used for patient controls.
3.
NOTE:
Westgard
®
rules are discussed in detail in
The Westgard
®
Rule selections are available on either of the
QC FILE SET UP
screens.