11-8
CELL-DYN
®
3200 System Operator’s Manual
9140181H—October 2001
Quality Control
Quality Control Menu
Section 11
4. The remainder of the screen displays the file information and the data. The
UPPER
and
LOWER LIMITS
and
TARGET MEAN
entered are displayed
immediately above the parameter names. The sequence number for each
result is displayed to the left of the data. The
DATE, TIME
and
OPERATOR
ID
when the sample was run are displayed to the right of the data.
5. The following QC Log codes are displayed in the column immediately
preceding the date. These codes are the same as those used in the Data Log:
O
— Sample was run in the Open mode
C
— Sample was run in the Closed mode
N
— Incomplete aspiration in the Open mode
I
— Incomplete aspiration in the Closed mode
K
— Flow Error occurred
Background and latex samples use only the O or C codes.
6. The statistics are displayed below the data as follows:
N:
the Number of runs used in the calculation
FILE MEAN:
the Mean value for the number of runs used in the
calculation
STD DEV:
the Standard Deviation (+/-1) for the number of runs used
in the calculation
CV (%):
the Coefficient of Variation in percent for the number of
runs used in the calculation
The following soft key labels are displayed on the
VIEW QC LOG
screen:
PURGE QC LOG
LEVEY-JENNINGS
REJECT SPECIMEN/
ACCEPT SPECIMEN
(Key label alternates between selections)
DELETE SPECIMEN
MOVE SPECIMEN
WRITE QC TO DISK
PRINT QC LOG
RETURN
The function of each of these keys is discussed in this section.