10-64
CELL-DYN
®
3200 System Operator’s Manual
9140181H—October 2001
Troubleshooting and Diagnostics
Tables for Instrument Conditions and Messages
Section 10
Loader Fault 45: unload zone full detector malfunction
/BL
The system failed to retract mechanical arms in unload zone.
– – – - Probable Cause – – –
– – – - Corrective Action –
1. A mechanical problem has
prevented the rack arms in the
unload zone from retracting.
1. Check for an obstruction that is
preventing the rack arms in the
unload zone from retracting and, if
one is found, remove it. Then, reset
the racks and Sample Loader
operation by pressing the following
keys in order:
[START LOADER]
,
[RESUME LOADER]
.
2. A failure of the sensor or sample
loader mechanics has occurred.
2. If the message appears repeatedly,
contact your Abbott Hematology
Customer Support Center.
Loader Fault 46: unload zone nearly full detector malfunction
/BL
The system is detecting an unload zone nearly full condition (4 racks in unload
zone) even though there are fewer than 4 racks in the unload zone.
– – – - Probable Cause – – –
– – – - Corrective Action –
1. A mechanical problem has
prevented the rack arms in the
unload zone from retracting.
1. Check for an obstruction that is
preventing the rack arms in the
unload zone from retracting and, if
one is found, remove it. Then, reset
the racks and Sample Loader
operation by pressing the following
keys in order:
[START LOADER]
,
[RESUME LOADER]
.
2. A failure of the sensor or sample
loader mechanics has occurred.
2. If the message appears repeatedly,
contact your Abbott Hematology
Customer Support Center.
Table 10.2 General Fault Conditions (Continued)