5-136
CELL-DYN
®
3200 System Operator’s Manual
9140181H—October 2001
Operating Instructions
Using The Data Log
Section 5
Transmit Data
The
[TRANSMIT DATA]
key is used to transmit a record to a Laboratory Information
System or on-line computer. When the
[TRANSMIT DATA]
key is pressed, the screen
prompts the operator to enter the starting and ending sequence numbers (from the
lowest to the highest) for the desired transmission. Records may be transmitted
singly or in batches as designated by the sequence number(s).
When the
[TRANSMIT DATA]
key is pressed, the <Starting Sequence #> field appears
in the upper left corner of the screen and the cursor is positioned in this field. The
operator should enter the sequence number of the first specimen to be transmitted
and press the Enter key. If the number is valid, the system accepts the entry and the
<Ending Sequence #> field appears in the upper left corner of the screen. The
operator should type in the ending sequence number and press Enter. The system
begins transmitting automatically. If only one record is to be transmitted, the
starting and ending sequence number will be the same.
NOTE:
Use the ESC key to cancel this function and return to the
DATA LOG
menu.
Use the Backspace key or left arrow key to cancel an entry and retype the
sequence number.
Because specimen records are shown in summary form on the
DATA LOG
menu,
only the summary data of these records will be transmitted. No histogram data
accompanies the summary data.
Print Data Log
The
[PRINT DATA LOG]
key is used to print the Data Log. When the
[PRINT DATA
LOG]
key is pressed, the screen prompts the operator to enter the starting and ending
sequence numbers (from the lowest to the highest) for the desired printout. (See
Figure 5.52.)
When the
[PRINT DATA LOG]
key is pressed, the <Starting Sequence #> field
appears in the upper left corner of the screen and the cursor is positioned in this
field. The operator should enter the sequence number of the first specimen to be
printed and press the Enter key. If the number is valid, the system accepts the entry
and the <Ending Sequence #> field appears in the upper left corner of the screen.
The operator should type in the ending sequence number and press Enter. The
system begins printing automatically. If only one record is to be printed, the starting
and ending sequence number will be the same.
NOTE:
Use the ESC key to cancel this function and return to the
DATA LOG
menu.
Use the Backspace key or left arrow key to cancel an entry and retype the
sequence number.
Because specimen records are shown in summary form on the
DATA LOG
menu,
only the summary data of these records will be printed. No histogram data
accompanies the summary data.
NOTE:
To print histogram data, the operator must press the
[DISPLAY SPECIMEN]
key in the
DATA LOG
menu to select and display an individual specimen.
Then, the
[PRINT REPORT]
key will print the results of the specimen
displayed on the screen, including histograms.