10-54
CELL-DYN
®
3200 System Operator’s Manual
9140181H—October 2001
Troubleshooting and Diagnostics
Tables for Instrument Conditions and Messages
Section 10
DMA controller error during list mode acquisition
/BL
NOTE:
DMA stands for Direct Memory Access.
A memory access control problem occurred during data acquisition.
– – – - Probable Cause – – –
– – – - Corrective Action –
1. A failure in the Analyzer hardware
has occurred.
1. Initialize the Analyzer by pressing
the
[INITIALIZATION]
key on the
second
DIAGNOSTICS MENU
screen. If the message appears
again, reboot the System.
NOTE:
If you wish to use the
Sample Loader, reset the
racks before initializing or
rebooting the System.
2. If the message appears repeatedly,
contact your Abbott Hematology
Customer Support Center.
DMA controller setup error
/BL
A problem occurred during setup of the memory access controller.
– – – - Probable Cause – – –
– – – - Corrective Action –
1. A failure in the Analyzer hardware
has occurred.
1. Initialize the Analyzer by pressing
the
[INITIALIZATION]
key on the
second
DIAGNOSTICS MENU
screen. If the message appears
again, reboot the System.
NOTE:
If you wish to use the
Sample Loader, reset the
racks before initializing or
rebooting the System.
2. If the message appears repeatedly,
contact your Abbott Hematology
Customer Support Center.
Table 10.2 General Fault Conditions (Continued)