10-12
CELL-DYN
®
3200 System Operator’s Manual
9140181H—October 2001
Troubleshooting and Diagnostics
Diagnostics Menu
Section 10
Figure 10.11 Pump Operation Screen — Vacuum ON
Vacuum On
When the
[VACUUM ON]
key is pressed, the key label changes to
[VACUUM OFF]
, the
vacuum pump is turned ON and the screen displays the message
Vacuum Is On
.
(See Figure 10.11.) Press the
[VACUUM OFF]
key to turn the pump OFF.
NOTE:
The pump is automatically turned OFF and control of the pump is
returned to the instrument when the screen is exited.
This key is useful for troubleshooting vacuum problems. If the pump does not turn
ON when the key is pressed, the pump may be the cause of the vacuum problem.
Pressure On
When the
[PRESSURE ON]
key is pressed, the key label changes to
[PRESSURE
OFF]
, the pressure pump is turned ON and the screen displays the message
Pressure Is On
. Press the
[PRESSURE OFF]
key to turn the pump OFF.
NOTE:
The pump is automatically turned OFF and control of the pump is
returned to the instrument when the screen is exited.
This key is useful for troubleshooting pressure problems. If the pump does not turn
ON when the key is pressed, the pump may be the cause of the pressure problem.