9-26
CELL-DYN
®
3200 System Operator’s Manual
9140181H—October 2001
Service and Maintenance
Monthly Maintenance Procedures
Section 9
6. Pinch off the tubing above the luer slip inlet with hemostats.
Clamp the
tubing above the Diluent/Sheath Filter.
NOTE:
Be sure not to damage the luer slip inlet port.
7. Disconnect the luer slip inlet from the used Diluent/Sheath Filter.
NOTE:
Use a paper towel to absorb any liquid discharged from the filter or
tubing.
8. Connect the luer slip inlet to the top of a new Diluent/Sheath Filter.
9. Disconnect the used Diluent/Sheath Filter from the silicon tubing at the lower
end and discard it.
10. Attach the silicon tubing to the lower part of the new Diluent/Sheath Filter.
11. Remove the hemostats.
12. Install the new Diluent/Sheath Filter into the spring clip.
13. Turn the Power Switch ON.
14. Wait until the instrument is initialized, then press
[RUN]
to prime the system
and place it in the READY state.
15. Run background cycles until the background reading is within specification,
usually five to ten cycles. Refer to Section 4 for the background specification.
16. Record the completion of this procedure in the maintenance log.