Calibration Procedures
Enter Factor Method
Section 6
Calculating New Closed Mode Calibration Factors
Based on the results from steps 1 through 5 above, perform the following:
1. If Closed Mode calibration is required, use Worksheet 6 to calculate the New
Closed Mode Calibration Factor for the parameters that require calibration.
(Use the Current Closed Mode Cal Factor that was printed in
'HWHUPLQLQJ
WKH&ORVHG0RGH0HDQ
step 3, at the start of this procedure.)
Calibrator Calibration:
Whole Blood Calibration:
For example, if the Reference Mean Value for WOC is 6.6, the CELL-DYN
Mean for WOC is 7.4, and the current Closed Mode Calibration Factor for
WOC is 1.04, then:
(6.6 / 7.4)
×
1.04 = 0.93
and 0.93 is the New Closed Mode Calibration Factor for WOC.
2a. If the new Closed Mode Calibration Factors falls within the acceptable range
given on the worksheet, follow the procedure for entering new calibration
factors below.
2b. If the New Closed Mode Calibration Factor for a given parameter falls
outside the acceptable range, there may be a computation error.
'RQRW
FDOLEUDWHWKDWSDUDPHWHU
Recheck all calculations. If the factor(s) is still
outside the acceptable range, call the Abbott Hematology Customer Support
Center for assistance (at 1-800-CELL-DYN in the U.S.).