5-22
CELL-DYN
®
3200 System Operator’s Manual
9140181H—October 2001
Operating Instructions
Set Up Instructions
Section 5
X-B Setup Menu
The
[X-B SET UP]
key is used to select the
X-B SET UP
screen. (See Figure 5.7.) This
screen is used to enter upper and lower acceptance limits, target values, and action
limits for the X-B Moving Average QC Program for both the WBC and RBC
parameters. The following soft key labels are displayed when the
[X-B SET UP]
key
is pressed:
TURN X-B RBC ON/
(The key alternates between the selections.)
TURN X-B RBC OFF
TURN X-B WBC ON/
(The key alternates between the selections.)
TURN X-B WBC OFF
PRINT
RETURN
The following options are available on the
X-B SET UP
screen shown in Figure 5.7:
1.
LOWER/UPPER LIMITS
The
Lower and Upper Limits
determine which patient results will be used
in the X-B Moving Average calculation. Results that fall outside these limits
are automatically excluded from the X-B calculations. These limits should be
set widely to exclude grossly abnormal samples (such as background counts)
that would bias the calculation but should include at least 95% of the patient
results.
2.
TARGET VALUE
The
Target Value
for X-B Analysis is similar to the assay value for a
commercial control. It is derived from the patient population that is analyzed
on the instrument.
3.
ACTION LIMIT
The
Action Limit
is the acceptable limit of variation around the target value.
NOTE:
The X-B Program is discussed in detail in
To Set Up X-B
1. Press
[X-B SET UP]
to display the
X-B SET UP
screen.
2. Use the Arrow keys on the keyboard to move the cursor to the desired entry
field.
3. Type the numbers and press the Enter key on the keyboard to save the entry
and advance the cursor to the next entry field.
4. Repeat steps 2 and 3 until all entries have been made.
5. If desired, press
[PRINT]
to obtain a printout of the entered values.
6. If the soft key
[TURN X-B RBC ON]
or
[TURN X-B WBC ON]
is displayed, press
the appropriate key to enable the X-B program.
NOTE:
For example, when the X-B RBC program is enabled the screen
displays the message
THE X-B RBC PROGRAM IS ON
and the
[TURN X-B RBC OFF]
key is displayed.
7. Press
[RETURN]
to return to the
QC SET UP MENU
screen.